A novel safety-enhanced optimal control algorithm for turbodynamic blood pumps successfully maintains cardiac perfusion while preventing suction events in simulations.
Answers are grounded in this paper and cite their sources.
Supports feasibility of suction-avoidant pump control in animals; leaves open clinical translation and outcome impact.
A safety-enhanced optimal (SEO) control algorithm for turbodynamic blood pump is proposed. Analysis of in vivo animal experimental data reveals that two new control indices-the gradient of pulsatility of pump pressure head with respect to pump speed and the gradient of minimum pump flow-have their peak within a proximity to the suction point but not at the exact suction point. They were also verified to satisfy the requirement of cost function for the extremum seeking control (ESC). New cost functions were tested for ESC to find and track the new operating point--SEO operating point--where sufficient cardiac perfusion and safety margin to suction is guaranteed. By computer simulation, it is confirmed that the SEO operating point was successfully found and tracked in both fixed and varying hemodynamic load scenarios using proposed control indices without resorting to a slope seeking control algorithm where the reference slope must be supplied.
No takes yet. Share an insight, caveat, or question.
Gwak et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: