Key result
The newly developed IoT-based Ankle Brachial Index measuring instrument, which simultaneously measures the right and left sides of the body, demonstrated an accuracy of 96.6% when compared to a standard sphygmomanometer.
Why the study?
Early detection of cardiovascular disease reduces mortality risk, but prior ABI tools could only measure one side of the body at a time.
A novel IoT-based device for simultaneous bilateral ABI measurement demonstrated high accuracy compared to standard manometry, offering a potential tool for efficient peripheral arterial disease screening.
Answers are grounded in this paper and cite their sources.
Simultaneous ABI device may aid early CVD detection in practice; leaves open need for clinical validation in prospective cohorts.
ABSTRAK Deteksi dini penyakit kardiovaskular sangat diperlukan untuk mengurangi risiko kematian. Deteksi dini penyakit kardiovaskular dapat dilakukan dengan bermacammacam metode, salah satunya adalah menggunakan metode Ankle Brachial Indeks (ABI). Metode ini membandingkan tekanan darah antara sistole pada bagian tangan dan kaki secara bersamaan. Pada penelitian ini dibuatlah alat pengukur ABI yang dapat mengukur secara serempak antara bagian tubuh kanan dan kiri, yaitu merupakan pengembangan dari penelitian sebelumnya yang hanya dapat melakukan pengukuran pada satu sisi tubuh saja. Dengan pengukuran secara serempak, diharapkan hasil yang diperoleh lebih akurat dan lebih efektif. Hasil validasi dari alat ini setelah dibandingkan dengan sphygmomanometer memiliki akurasi sebesar 96.6%. Selain itu data riwayat pemeriksaan dapat disimpan dan diakses oleh pasien dan dokter melalui teknologi IoT. Kata kunci: deteksi dini, kardiovaskular, Ankle Brachial Indeks, IoT ABSTRACT Early detection of Cardiovascular Disease (CVD) is needed to reduce the risk of death. Early detection of cardiovascular disease can be done using various methods, one of which is the Ankle Brachial Index (ABI) method. This method compares blood pressure between systoles on the hands and feet simultaneously. In this study, the ABI measuring instrument was made that could simultaneously measure the right and left parts of the body, a development from previous research that could only take measurements on one side of the body. With simultaneous measurements, the results will be more accurate and effective. The validation results of this tool, when compared with the sphygmomanometer, have an accuracy of 96.6%. Besides, patients and doctors can store and access examination history data through IoT platform.
No takes yet. Share an insight, caveat, or question.
Dewi et al. (2023) studied Healthy (Device Validation for Peripheral Arterial Disease) (n=3). IoT-based simultaneous Ankle Brachial Index (ABI) measurement tool vs. Aneroid sphygmomanometer was evaluated on Accuracy of the ABI measuring instrument compared to a standard sphygmomanometer. The newly developed IoT-based Ankle Brachial Index measuring instrument, which simultaneously measures the right and left sides of the body, demonstrated an accuracy of 96.6% when compared to a standard sphygmomanometer.
Synapse has enriched 10 related papers. Consider them for comparative context: