Key result
The Alere i Influenza A&B assay demonstrated sensitivities of 65.96% and 53.33% for influenza A and B, respectively, and specificities of 98.51% and 95.96% compared to molecular techniques.
Why the study?
Does the Alere i Influenza A&B assay accurately detect influenza A and B viruses compared to molecular techniques in inpatients with influenza-like illness?
Observational (n=119)
Does the Alere i Influenza A&B assay accurately detect influenza A and B viruses compared to molecular techniques in inpatients with influenza-like illness?
The Alere i Influenza A&B assay is more sensitive than the BinaxNOW card but has moderate sensitivity compared to gold-standard molecular techniques, while maintaining excellent specificity.
Answers are grounded in this paper and cite their sources.
Moderate sensitivity cautions against sole use to rule out influenza in inpatients; leaves open need for prospective validation versus molecular standards.
The Alere i Influenza A&B assay is a novel isothermal nucleic acid amplification assay capable of detecting and differentiating influenza A and B viruses in approximately 15 min with minimal hands-on time. This study was conducted in order to assess the performance of the Alere i Influenza A&B assay compared to molecular techniques, considered to be gold standard methods, to evaluate the results. A total of 119 nasopharyngeal swabs collected from inpatients with influenza-like illness were included in the study using both archived and prospectively collected samples from adults and children. Prospectively collected samples were also compared to the Alere BinaxNOW® Influenza A & B Card. The overall sensitivity for detection of influenza A and B viruses compared to those of molecular techniques were 65.96 % and 53.33 % respectively, while the specificity was 98.51 % and 95.96 %. Compared to the Alere BinaxNOW® Influenza A & B Card, the Alere i assay is considerably more sensitive for detection of influenza A and B viruses, although both tests demonstrated excellent specificity for diagnosis of influenza viruses.
No takes yet. Share an insight, caveat, or question.
Chiarella et al. (2016) conducted an observational in influenza-like illness (n=119). Alere i Influenza A&B assay vs. Molecular techniques and Alere BinaxNOW Influenza A & B Card was evaluated on Detection of influenza A and B viruses. The Alere i Influenza A&B assay demonstrated sensitivities of 65.96% and 53.33% for influenza A and B, respectively, and specificities of 98.51% and 95.96% compared to molecular techniques.
Synapse has enriched 10 related papers. Consider them for comparative context: