Key result
An optimized new generation fluoroscopy system resulted in lower radiation doses across all electrophysiology procedure groups compared to an older system.
Why the study?
Does an optimized new generation fluoroscopy system reduce radiation dose in patients undergoing electrophysiology procedures compared to an older system?
Cohort
No
Does an optimized new generation fluoroscopy system reduce radiation dose in patients undergoing electrophysiology procedures compared to an older system?
Optimized contemporary digital fluoroscopy systems can achieve ultra-low radiation levels for electrophysiology procedures without compromising procedural time or success.
Answers are grounded in this paper and cite their sources.
Supports lower radiation with new fluoroscopy in EP procedures; extends observational data but leaves open randomized confirmation.
BACKGROUND: Electrophysiology procedures require fluoroscopic guidance, with the associated potentially adverse effects of ionizing radiation. Newer fluoroscopy systems have more features that enable dose-reduction strategies. This study aimed to investigate any reduction in radiation dose between an older fluoroscopy system (Philips Integris H5000, Philips Healthcare, Einhoven, Netherlands) and one of the latest systems (Siemens Artis Q, Siemens Healthcare, Erlangen, Germany), optimized with dose-reduction strategies. METHODS: Radiation dose measures were collected over a 2-year period in a single electrophysiology laboratory. Procedures were separated into seven groups: devices, biventricular devices, electrophysiology studies, standard radiofrequency ablation, complex atrial ablation, ablation for ventricular arrhythmias, and pulmonary vein isolation. In the first year, an older fluoroscopy system was used, and in the second year, a new system, with dose reduction strategies. Comparisons were also made to the literature with regard to radiation dose levels. RESULTS: [P > 0.001]) with the new system and was lower across all groups. DAP readings with the new system are some of the lowest published in the literature in all groups. CONCLUSION: An optimized contemporary digital fluoroscopy system, with low radiation dose configuration and continued good procedural practice, can result in ultra-low radiation levels for all electrophysiology procedures, without compromising procedural time or procedural success.
No takes yet. Share an insight, caveat, or question.
Crowhurst et al. (2017) conducted a cohort in Electrophysiology procedures. Optimized new generation fluoroscopy system (Siemens Artis Q) vs. Older fluoroscopy system (Philips Integris H5000) was evaluated on Radiation dose (DAP readings) (p=> 0.001). An optimized new generation fluoroscopy system resulted in lower radiation doses across all electrophysiology procedure groups compared to an older system.