Key result
Photoplethysmographic pulse variation >9% identified patients with arterial pulse pressure variation >13%, serving as a noninvasive predictor of fluid responsiveness.
Why the study?
Can photoplethysmographic pulse variation predict fluid responsiveness similarly to arterial pulse pressure variation in mechanically ventilated patients?
Observational (n=57)
Can photoplethysmographic pulse variation predict fluid responsiveness similarly to arterial pulse pressure variation in mechanically ventilated patients?
Answers are grounded in this paper and cite their sources.
Photoplethysmography can serve as a noninvasive predictor of fluid responsiveness in mechanically ventilated patients, correlating well with arterial pulse pressure variation.
In Brief We analyzed ventilation-induced changes in arterial blood pressure and photoplethysmography from waveforms obtained by monitoring 57 patients in the operating room and intensive care unit. Analysis of systolic and pulse pressure variations during positive pressure ventilation, ΔUp, ΔDown, and changes in the preejection period on both arterial and photoplethysmographic waveforms were possible in 49 (86%) patients. The pulse pressure variation and preejection period were similar when calculated using both arterial blood pressure and photoplethysmography, whereas the other variables were different. Photoplethysmographic pulse variation >9% identified patients with arterial pulse pressure variation >13% (area under ROC curve = 0.85) or ΔDown >5 mm Hg (area under ROC curve = 0.85). In hypotensive patients, photoplethysmographic pulse variation >9% remained the best threshold value (pulse pressure variation >13%: area under ROC curve = 0.90; ΔDown >5 mm Hg: area under ROC curve = 0.93) for predicting fluid responsiveness. In conclusion, this study showed that pulse variations observed in the arterial pressure waveform and photoplethysmogram are similiar in response to positive pressure ventilation. Furthermore, photoplethysmographic pulse variation > 9% identifies patients with ventilation-induced arterial blood pressure variation that is likely to respond to fluid administration. IMPLICATIONS: Photoplethysmographic heart rate variation in response to positive pressure ventilation was related to pulse pressure variation in the arterial pressure waveform. Because pulse pressure variation can predict fluid responsiveness, photoplethysmography could be used as a noninvasive predictor of fluid responsiveness.
No takes yet. Share an insight, caveat, or question.
Natalini et al. (2006) conducted an observational in Mechanical ventilation (n=57). Photoplethysmography vs. Arterial blood pressure waveform was evaluated on Identification of arterial pulse pressure variation >13% (AUC 0.85). Photoplethysmographic pulse variation >9% identified patients with arterial pulse pressure variation >13%, serving as a noninvasive predictor of fluid responsiveness.
Synapse has enriched 10 related papers. Consider them for comparative context: