Answers are grounded in this paper and cite their sources.
A transmission x-ray imaging setup based on a confocal combination of a polycapillary focusing x-ray optic followed by a polycapillary collimating x-ray optic was designed and demonstrated to have good resolution, better than the unmagnified pixel size and unlimited by the x-ray tube spot size. This imaging setup has potential application in x-ray imaging for small samples, for example, for histology specimens.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: