Why the study?
Watchman device buckling is a previously undescribed deployment complication.
Can intraprocedural TEE detect and correct Watchman device buckling during LAA occlusion?
Can intraprocedural TEE detect and correct Watchman device buckling during LAA occlusion?
Intraprocedural 2D and 3D TEE is crucial for identifying and correcting Watchman device buckling, a newly described complication during LAA occlusion.
Answers are grounded in this paper and cite their sources.
Highlights buckling as a TEE-detectable Watchman complication in difficult LAA anatomy; leaves open its incidence and optimal prevention.
• Watchman device buckling is a previously undescribed deployment complication. • Both 2D and 3D TEE assist in detection of buckling during and after the procedure. • Challenging LAA anatomy may play a significant role in buckling risk. • An acute angle between the deployment catheter and the LAA ostium may confer risk. • With diligent TEE guidance, buckling may be corrected during the procedure.