Key result
A 22 mm Amulet device was successfully deployed to completely seal a peri-device leak in a patient who previously received a Watchman 2.5 device.
Why the study?
Peri-device leak following left atrial appendage closure is often treated with coils, plugs, or second occluders, but use of the Amulet device had not been reported.
Case Report (n=1)
No
This case report provides the first evidence that an Amulet device can be successfully used to seal a peri-device leak following incomplete left atrial appendage closure with a Watchman device.
May serve as bailout for Watchman peri-device leak; hypothesis-generating and requires confirmation in larger studies.
Peri-device leak after left atrial appendage closure (LAAC) is often treated with endovascular coils, plugs, or second occluders. This is the first study reporting the Amulet device used for peri-device leak. An 80-year-old male with paroxysmal atrial fibrillation and recurrent falls with head trauma who underwent LAAC with a 24 mm Watchman 2.5 device 3 years ago at another institution was referred to our clinic for management of the peri-device leak. Transesophageal echocardiogram showed persistent residual peri-device leak with 5 mm width along the Coumadin ridge aspect of the device. Computed tomography (CT) also showed the peri-device leak with width of 6 mm and complete opacification of left atrial appendage (LAA). Importantly, CT demonstrated that the Watchman 2.5 device was deployed at distal LAA, leaving the proximal part of LAA with length of 10 mm from ostium. Under general anesthesia, a 22 mm Amulet device was deployed successfully with complete sealing of LAA. Procedure planning is the key to minimize the risk of peri-device leak or device-related thrombosis. Careful assessment of LAA anatomy using multimodality images for peri-device leak after LAAC helped optimal treatment strategy including second LAAC with different type of devices.
No takes yet. Share an insight, caveat, or question.
Yoon et al. (2023) conducted a case report in Peri-device leak after left atrial appendage closure (n=1). Amulet device implantation was evaluated on Complete sealing of LAA. A 22 mm Amulet device was successfully deployed to completely seal a peri-device leak in a patient who previously received a Watchman 2.5 device.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: