Key result
Computerized processing of video image series for real-time digitization of arteriograms yielded flow distribution values at the renal and iliac arteries accurate within +/- 5% of total abdominal aortic flow.
Why the study?
Does computerized processing of digital angiography accurately measure regional arterial blood flow in a porcine model?
Does computerized processing of digital angiography accurately measure regional arterial blood flow in a porcine model?
Computerized processing of digital angiography can accurately measure regional arterial blood flow in a preclinical model, with potential applications in cerebral, pulmonary, and coronary arteries.
Answers are grounded in this paper and cite their sources.
Accurate in porcine arteriography; extends validation of digital flow quantification but leaves open human translation.
A system for computerized processing of video image series was utilized for the real-time digitization of arteriograms. There were 256 x 256 picture elements (pixels) per image with a resolution of 256 gray levels. Temporal changes in x-ray absorption at each pixel as induced by contrast media were computed for the extraction of regional blood flow parameters. Four steps of processing included: (a) vascular masking, (b) time parameter extraction, (c) segmentation of vessels, and (d) segmental volume and flow determinations. The results were displayed pictorially as angiographic parameter images. Measurement of the regional blood flow was tested at the abdominal arteries in nine pigs. Flow distribution values at the renal and iliac arteries were accurate within +/- 5% of the total abdominal aortic flow. The method may also be applicable to quantitative circulatory studies of the cerebral, pulmonary, and possibly, coronary arteries.
No takes yet. Share an insight, caveat, or question.
Bürsch et al. (1981) studied this question. Computerized processing of video image series for real-time digitization of arteriograms was evaluated on Flow distribution values at the renal and iliac arteries. Computerized processing of video image series for real-time digitization of arteriograms yielded flow distribution values at the renal and iliac arteries accurate within +/- 5% of total abdominal aortic flow.
Synapse has enriched 10 related papers. Consider them for comparative context: