Key result
Electromagnetic flow velocity catheters and a new valve insufficiency analyser successfully evaluated regurgitant flow through competent and incompetent pulmonary valves in animal experiments.
Answers are grounded in this paper and cite their sources.
Electromagnetic flow velocity catheters combined with a new analog computing circuit can accurately evaluate regurgitant flow in pulmonary valves in animal models.
In animal experiments regurgitant flow through competent and incompetent pulmonary valves was evaluated with electromagnetic flow velocity catheters by using the ratio of the separately integrated backward and forward flow velocity signals. This ratio was obtained with a special purpose analog computing circuit set by an in-vitro zero flow reference signal, which was in good agreement with in-vivo zero flow signals.
No takes yet. Share an insight, caveat, or question.
Rohmer et al. (1976) studied Pulmonary valve incompetence. Electromagnetic flow velocity catheters and a new valve insufficiency analyser was evaluated on Regurgitant flow ratio. Electromagnetic flow velocity catheters and a new valve insufficiency analyser successfully evaluated regurgitant flow through competent and incompetent pulmonary valves in animal experiments.
Synapse has enriched 10 related papers. Consider them for comparative context: