Key result
The AMPLICOR Enterovirus Test demonstrated a sensitivity of 96.3% and a specificity of 100% compared with culture for detecting enterovirus in cerebrospinal fluid specimens.
Why the study?
Does the AMPLICOR Enterovirus Test improve the detection of enterovirus infection in CSF specimens compared to viral culture?
Does the AMPLICOR Enterovirus Test improve the detection of enterovirus infection in CSF specimens compared to viral culture?
The AMPLICOR Enterovirus Test provides rapid and highly sensitive detection of enterovirus RNA in CSF compared to traditional viral culture.
Answers are grounded in this paper and cite their sources.
May support rapid PCR testing over culture for enteroviral meningitis; leaves open prospective validation before routine use.
The AMPLICOR Enterovirus Test was evaluated with 103 cerebrospinal fluid (CSF) specimens. Twenty-seven CSF specimens were culture positive. With the AMPLICOR test, enterovirus RNA was detected in 34 specimens. Compared with culture, the AMPLICOR test gave a sensitivity of 96.3% and a specificity of 100%. The sensitivity of culture was 79.4% in comparison with the AMPLICOR test.
No takes yet. Share an insight, caveat, or question.
Kessler et al. (1997) studied this question. The AMPLICOR Enterovirus Test demonstrated a sensitivity of 96.3% and a specificity of 100% compared with culture for detecting enterovirus in cerebrospinal fluid specimens.
Synapse has enriched 10 related papers. Consider them for comparative context: