Key result
Treatment of falsely elevated AxSYM cardiac troponin I samples with heterophile-blocking reagents paradoxically increased the magnitude of the false results 17- and 26-fold in a 53-year-old woman.
Why the study?
What is the nature of the interfering substance causing false-positive AxSYM cardiac troponin I results in this patient, and how can it be neutralized?
Case Report (n=1)
What is the nature of the interfering substance causing false-positive AxSYM cardiac troponin I results in this patient, and how can it be neutralized?
Clinicians and laboratorians must be aware of atypical immunoassay interferences causing false-positive troponin results that may not be neutralized by standard heterophile-blocking reagents.
Clinicians should suspect atypical interferences in unexplained false-positive troponin results; leaves open characterization of non-heterophile blockers for immunoassay artifacts.
A number of classes of endogenous antibodies, including heterophile, rheumatoid factor, and autoantibodies, can interfere with immunoassay measurements of many different analytes. Heterophile and rheumatoid factor antibody interferences have been described previously for the AxSYM cardiac troponin I assay. Several commercial products have been developed to neutralize heterophile antibody interferences. We describe a patient with multiple apparently falsely elevated cardiac troponin I results that were unique to the AxSYM analyzer. These cardiac troponin I results diluted linearly. When treated with 2 different heterophile-blocking reagents, the magnitudes of the falsely elevated results increased 17- and 26-fold, and these results also demonstrated dilution linearity. This interfering substance could be removed by passage through an immobilized protein A column and by polyethylene glycol precipitation. It does not appear to be a classic heterophile antibody, nor is it a paraprotein. Laboratorians must remain constantly vigilant for immunoassay interferences that lead to clinically significant inaccurate results and must recognize that accepted methods for detecting and neutralizing the interference may be ineffective.
No takes yet. Share an insight, caveat, or question.
Knoblock et al. (2002) conducted a case report in False-positive cardiac troponin I (n=1). Heterophile-blocking reagents was evaluated on Magnitude of falsely elevated results after treatment with heterophile-blocking reagents. Treatment of falsely elevated AxSYM cardiac troponin I samples with heterophile-blocking reagents paradoxically increased the magnitude of the false results 17- and 26-fold in a 53-year-old woman.