Key result
Polytetrafluoroethylene (PTFE) suture exhibited a significantly higher elastic modulus than porcine mitral valve chordae, though it demonstrated some viscoelastic characteristics approaching natural chordal behavior.
Why the study?
Does PTFE suture simulate the mechanical function of porcine mitral valve chordae tendineae?
Does PTFE suture simulate the mechanical function of porcine mitral valve chordae tendineae?
PTFE suture exhibits some viscoelastic characteristics approaching natural chordae tendineae, making it the best current synthetic alternative for chordal replacement despite having a significantly higher elastic modulus.
Supports PTFE as current chordal replacement option; leaves open refinement of materials to match native elastic modulus.
Recent reports have advocated the use of polytetrafluoroethylene (PTFE) suture for replacement or reinforcement of ruptured or elongated mitral valve chordae tendineae. The mechanical properties of PTFE (Gore-Tex) and other sutures were determined and compared to those of porcine mitral valve chordae. The results were analyzed to assess how closely chordal mechanical function may be simulated by synthetic suture materials. Chordae tendineae and suture samples were tested in uniaxial tension using an INSTRON Model 1000 at strain rates of 5 and 10 mm/min. The stress (g/mm2) was plotted versus strain, and the elastic modulus determined as the slope of the curve. Chordae tendineae exhibited a nonlinear viscoelastic stress/strain behavior. The elastic modulus of both suture types tested was significantly higher than that of the chordae. However, the PTFE suture did exhibit some viscoelastic characteristics (hysteresis and creep) that begin to approach the chordal behavior. Chordal viscoelastic behavior results from the inherent composite structure (collagen, elastin, endothelium, water, and ground substance). As yet, no synthetic materials are able to imitate this behavior with the appropriate tensile strength and fatigue resistant characteristics. At present, PTFE appears to be the best synthetic alternative for chordal replacement, due to its limited viscoelastic capabilities. Nevertheless, the need to more nearly approximate the mechanical behavior of mitral valve chordae tendineae with synthetic material warrants further investigation.
No takes yet. Share an insight, caveat, or question.
Cochran et al. (1991) studied Ruptured or elongated mitral valve chordae tendineae. Polytetrafluoroethylene (PTFE) suture vs. Porcine mitral valve chordae tendineae was evaluated on Elastic modulus and viscoelastic stress/strain behavior. Polytetrafluoroethylene (PTFE) suture exhibited a significantly higher elastic modulus than porcine mitral valve chordae, though it demonstrated some viscoelastic characteristics approaching natural chordal behavior.
Synapse has enriched one closely related paper. Consider it for comparative context: