Key result
A dedicated radiofrequency guidewire required significantly fewer attempts to achieve transseptal puncture in porcine hearts compared to electrified 0.014” (1.10 vs 2.17; p<0.01) or 0.032” guidewires.
Why the study?
Electrocautery applied to metal guidewires can achieve transseptal puncture, but dedicated radiofrequency guidewires may represent a better alternative.
Does a dedicated radiofrequency guidewire reduce the number of attempts required to achieve transseptal puncture compared to electrified metal guidewires in porcine hearts?
Does a dedicated radiofrequency guidewire reduce the number of attempts required to achieve transseptal puncture compared to electrified metal guidewires in porcine hearts?
Absolute Event Rate: 1.1% vs 2.17%
p-value: p=<0.01
Dedicated RF guidewires require fewer attempts and cause less tissue damage and heating compared to electrified metal guidewires for transseptal puncture in a porcine model.
Supports dedicated RF wires for fewer transseptal attempts in animals; leaves open human efficacy and safety.
Application of electrocautery to a metal guidewire can be used to perform transseptal puncture (TSP). Dedicated radiofrequency guidewires (RF) may represent a better alternative. This study compares safety and effectiveness of electrified guidewires to a dedicated RF wire. TSP was performed on porcine hearts using an electrified 0.014” or 0.032” guidewire under various power settings compared to TSP using a dedicated RF wire with 5W power. The primary endpoint was the number of attempts required to achieve TSP. Secondary endpoints included the rate of TSP failure, TSP consistency, effect of the distance between tip of the guidewire and the tip of the dilator, and effect of RF power output level. Qualitative secondary endpoints included tissue puncture defect appearance, thermal damage to the TSP guidewire or dilator, and tissue temperature using thermal imaging. The RF wire required 1.10 ± 0.47 attempts to cross the septum. The 0.014” electrified guidewire required 2.17 ± 2.36 attempts (2.0x higher than the RF wire; p<0.01), and the 0.032” electrified guidewire required 3.90 ± 2.93 attempts (3.5x higher than the RF wire; p<0.01). Electrified guidewires had a higher rate of TSP failure, larger defects, more tissue charring, higher temperatures, and greater tissue heating. Fewer RF applications were required to achieve TSP using a dedicated RF wire compared to an electrified guidewire. Smaller defects and lower tissue temperatures were also observed using the RF wire. Electrified guidewires required greater energy delivery and were associated with equipment damage and tissue charring.
No takes yet. Share an insight, caveat, or question.
Wasserlauf et al. (2021) studied Transseptal puncture. Dedicated radiofrequency (RF) guidewire vs. Electrified 0.014” or 0.032” metal guidewire was evaluated on Number of attempts required to achieve transseptal puncture (p=<0.01). A dedicated radiofrequency guidewire required significantly fewer attempts to achieve transseptal puncture in porcine hearts compared to electrified 0.014” (1.10 vs 2.17; p<0.01) or 0.032” guidewires.