Key result
Electrified guidewire TSP via PFA sheath is linked to ~11% faster punctures versus traditional techniques.
Why the study?
Transseptal puncture applying electrocautery to a guidewire directly via a large pulsed field ablation sheath had not been described for left atrial access.
Does an electrified guidewire transseptal puncture via a large PFA sheath improve procedural efficiency and safety compared to traditional needle transseptal puncture in patients undergoing left atrial ablation?
Cohort (n=98)
Does an electrified guidewire transseptal puncture via a large PFA sheath improve procedural efficiency and safety compared to traditional needle transseptal puncture in patients undergoing left atrial ablation?
Absolute Event Rate: 16% vs 18%
p-value: p=0.024
Transseptal puncture using an electrified guidewire directly through a large pulsed field ablation sheath is a safe, effective, and efficient zero-exchange approach for left atrial access.
May streamline transseptal access; hypothesis-generating and requires randomized trials before practice change.
Background Transseptal puncture (TSP) with application of electrocautery to a guidewire has not been described directly via a large, pulsed field ablation (PFA) sheath to facilitate left atrial (LA) access. Objectives : To describe and characterize the efficacy and safety of our electrified 0.035" guidewire TSP (ETSP) technique via a 17Fr Faradrive PFA sheath. Methods We analyzed success, TSP time (from groin puncture to first LA electrogram), and complications of ETSP with a standard 0.035" Rosen guidewire directly via a Faradrive sheath. These were compared to patients undergoing left atrial ablation via: 1) traditional TSP with a Swartz Left (SL) sheath/BRK needle (non-ETSP) for Farapulse PFA; and 2) ETSP with smaller, non-Faradrive sheaths for radiofrequency ablation (RFA). Results A total of 98 patients were included: 58 ETSP with Faradrive sheath, 29 non-ETSP with SL sheath for subsequent PFA, and 11 ETSP with non-Faradrive sheath for RFA. ETSP directly via a Faradrive sheath was successful in 54 (93%) and 58 (100%) of patients after the first and second attempt respectively. No equipment exchange was required with ETSP in contrast to at least 3 with non-ETSP. There was no periprocedural stroke, cardiac tamponade or systemic embolism following ETSP. The median TSP time was significantly shorter with an ETSP compared to a non-ETSP technique (16min [15-18] vs 18min [16-20]p=0.024). Conclusions TSP with an electrified 0.035" guidewire via a large PFA sheath is effective and safe. This zero-exchange approach is more efficient than a traditional needle TSP combination.
No takes yet. Share an insight, caveat, or question.
Valappil et al. (2026) conducted a cohort in Patients undergoing left atrial ablation (n=98). Electrified 0.035" guidewire transseptal puncture (ETSP) via a 17Fr Faradrive PFA sheath vs. Traditional TSP with a Swartz Left sheath/BRK needle and ETSP with smaller non-Faradrive sheaths was evaluated on TSP time (from groin puncture to first LA electrogram) (p=0.024). Electrified guidewire transseptal puncture via a large pulsed field ablation sheath significantly reduced median transseptal puncture time compared to traditional techniques (16 vs 18 min; p=0.024).