Key result
Direct transseptal puncture with an RF guidewire and 16.8F PFA sheath achieves 100% complication-free success.
Why the study?
PFA catheter delivery typically requires standard transseptal puncture and an over-the-wire sheath exchange, which limits workflow and lengthens procedure time.
Is direct transseptal puncture using a radiofrequency guidewire combined with a 16.8F pulsed field ablation sheath safe and feasible in patients undergoing atrial fibrillation ablation?
Observational (n=166)
No
Is direct transseptal puncture using a radiofrequency guidewire combined with a 16.8F pulsed field ablation sheath safe and feasible in patients undergoing atrial fibrillation ablation?
A simplified direct transseptal puncture workflow using a radiofrequency guidewire and a 16.8F pulsed field ablation sheath is highly feasible and safe, eliminating the need for over-the-wire sheath exchanges.
Supports feasibility of simplified TSP in PFA ablation; leaves open need for prospective validation before practice change.
Transseptal puncture (TSP) is widely used in catheter-based cardiac interventions to gain left atrial (LA) access, but its workflow has remained largely unchanged and is still a source of serious complications. Pulsed field ablation (PFA) for pulmonary vein isolation (PVI) has been shown to be at least comparable with radiofrequency ablation (RFA) in terms of safety and efficacy. However, PFA catheter delivery to the LA typically requires a standard TSP and an over the wire sheath exchange which can limit workflow and lengthen procedure time-a shorter procedure time being a proposed advantage of PFA over RFA. This study aimed to evaluate a simplified workflow for direct TSP using the PFA sheath (Faradrive, Boston Scientific). We retrospectively analyzed 166 patients undergoing PVI with PFA in our center. TSP was performed by combining a 16.8F PFA sheath (Faradrive, Boston Scientific) and a RF-guidewire (Versacross, Boston Scientific) as a direct approach without the need over an over-the-wire exchange. The patient population had a mean age of 63.8 ± 11.3 years and was 72.3% male (n = 119/166). TSP was achieved in all patients (n = 166, 100%) without complication. The mean time from groin puncture to LA ablation catheter delivery was 16.2 ± 5.5 min with a mean fluoroscopy time of 15.7 ± 12.7 min. Dilator and sheath crossed the septum with no significant resistance in all cases (n = 168, 100%). The RF-guidewire allowed catheters to be tracked back up to the superior vena cava without exchange in cases where another dropdown was desired to locate a preferred puncture site. The stiffness of the wire provided adequate support to advance all sheaths to the left side regardless of outer diameter. This is the first case series on the use of a RF-guidewire combined with the PFA sheath for TSP. This study proved that an over a RF-powered guidewire TSP directly with 16.8F PFA sheath is feasible, reproducible, and safe. This very simplified workflow eliminates the need for both a rigid metal needle and an over the wire sheath exchange reducing procedure time and complexity, fluoroscopy time and potential related risks.
No takes yet. Share an insight, caveat, or question.
Jauvert et al. (2025) conducted an observational in Atrial Fibrillation (n=166). Direct transseptal puncture using a radiofrequency guidewire combined with a 16.8F pulsed field ablation sheath was evaluated on Successful transseptal puncture without complication. Direct transseptal puncture using a radiofrequency guidewire combined with a 16.8F pulsed field ablation sheath was successfully achieved in 100% of patients without any complications.