Key result
Inert gas rebreathing provided repeatable measurements of cardiac output and demonstrated a significant age-related decline in cardiac output in both males and females.
Why the study?
The study evaluated an inert gas rebreathing method for measuring cardiac output and aimed to provide robust normative data describing the influence of age, gender, and body size.
Does inert gas rebreathing using the Innocor device provide repeatable and physiologically sensitive measurements of cardiac output in healthy adults?
Cross-Sectional (n=1,505)
No
Does inert gas rebreathing using the Innocor device provide repeatable and physiologically sensitive measurements of cardiac output in healthy adults?
Inert gas rebreathing is a repeatable, non-invasive method for measuring cardiac output that is sensitive to acute physiological changes and demonstrates a significant age-related decline in cardiac output independent of body size.
Answers are grounded in this paper and cite their sources.
Supports repeatable non-invasive cardiac output assessment in research; leaves clinical adoption and age-related monitoring open pending prospective validation.
The aim of this study was to evaluate an inert gas rebreathing method (Innocor) for measurement of cardiac output and related haemodynamic variables and to provide robust normative data describing the influence of age, gender and body size on these variables. Four separate studies were conducted: measurement repeatability (study 1, n = 45); postural change (study 2, n = 40); response to submaximal cycling exercise (study 3, n = 20); and the influence of age, gender and body size (study 4, n = 1400). Repeated measurements of cardiac output, stroke volume and heart rate were similar, with low mean (±SD) differences (0.26 ± 0.53 L/min, 0 ± 11 mL and 2 ± 6beats/min, respectively). In addition, cardiac output and stroke volume both declined progressively from supine to seated and standing positions (P < 0.001 for both) and there was a stepwise increase in both parameters moving from rest to submaximal exercise (P < 0.001 for both). In study 4, there was a significant age-related decline in cardiac output and stroke volume in males and females, which remained significant after adjusting for body surface area (BSA, P < 0.001 for all comparisons). Both parameters were also significantly higher in those with high body mass index (BMI; P < 0.01 versus those with normal BMI for all comparisons), although indexing cardiac output and stroke volume to BSA reversed these trends. Inert gas rebreathing using the Innocor device provides repeatable measurements of cardiac output and related indices, which are sensitive to the effects of acute physiological manoeuvres. Moreover, inert gas rebreathing is a suitable technique for examining chronic influences such as age, gender and body size on key haemodynamic components of the arterial blood pressure.
No takes yet. Share an insight, caveat, or question.
Middlemiss et al. (2018) conducted a cross-sectional in Healthy adults (n=1,505). Inert gas rebreathing was evaluated on Mean difference between repeated measurements of cardiac output (0.26 L/min). Inert gas rebreathing provided repeatable measurements of cardiac output and demonstrated a significant age-related decline in cardiac output in both males and females.