Key result
TAVI with a balloon-expandable valve was systematically associated with acute stent recoil, with minimal lumen diameter decreasing by 1.10 mm after balloon deflation.
Observational (n=111)
TAVI with balloon-expandable valves is systematically associated with acute stent recoil, which is predicted by valve oversizing and prosthesis type but does not significantly impact haemodynamic performance.
Answers are grounded in this paper and cite their sources.
Acute recoil with balloon-expandable TAVI should not alter valve selection; leaves open long-term durability effects and optimization needs.
AIMS: The elastic behaviour (acute recoil) of a valve prosthesis stent following transcatheter aortic valve implantation (TAVI) is unknown. This study sought to determine the occurrence, severity, predictive factors and haemodynamic consequences of acute recoil following TAVI. METHODS AND RESULTS: A prospective angiographic analysis of the stent frame dimensions in 111 consecutive patients who underwent TAVI with a balloon-expandable valve (36 Edwards SAPIEN; 75 SAPIEN XT) was performed. Acute recoil was defined as the difference between minimal lumen diameter (MLD) at full balloon expansion and immediately after balloon deflation. MLD during balloon inflation was significantly larger than MLD after balloon deflation (23.40±2.31 mm vs. 22.29±2.21 mm, p<0.001), which represented an absolute and percent decrease in stent dimension of 1.10±0.40 mm and 4.70±1.76%, respectively. In the multivariate analysis, the predictors of larger recoil were a higher prosthesis/annulus ratio (r²=0.0624, p=0.015) and the SAPIEN XT prosthesis (r²=0.1276, p=0.001). No significant changes in haemodynamic performance were observed at discharge and follow-up in patients with larger recoil. CONCLUSIONS: TAVI with a balloon-expandable valve was systematically associated with a certain degree of valve stent recoil after balloon deflation. A higher degree of valve oversizing and the SAPIEN XT prosthesis predicted a larger degree of stent recoil.
No takes yet. Share an insight, caveat, or question.
Nombela‐Franco et al. (2014) conducted an observational in Patients undergoing TAVI with a balloon-expandable valve (n=111). TAVI with a balloon-expandable valve vs. Full balloon expansion vs immediately after balloon deflation was evaluated on Minimal lumen diameter (MLD) at full balloon expansion vs immediately after balloon deflation (MD 1.10 mm, p=<0.001). TAVI with a balloon-expandable valve was systematically associated with acute stent recoil, with minimal lumen diameter decreasing by 1.10 mm after balloon deflation.
Synapse has enriched 10 related papers. Consider them for comparative context: