Percutaneous pull-down maneuvers successfully repositioned malpositioned or dislocated aortic endografts in all 7 cases (100%) with no procedure-related complications.
Why the study?
Do percutaneous pull-down maneuvers safely reposition malpositioned or dislocated aortic endoprostheses?
Case Report (n=7)
Do percutaneous pull-down maneuvers safely reposition malpositioned or dislocated aortic endoprostheses?
Percutaneous pull-down maneuvers are technically feasible for repositioning misplaced or dislocated aortic endoprostheses without immediate complications.
Answers are grounded in this paper and cite their sources.
May support percutaneous repositioning in select cases; leaves open need for prospective validation of safety and durability.
PURPOSE: To present the capabilities and potential complications of 2 percutaneous techniques for repositioning malpositioned or dislodged aortic endografts. METHODS: Seven male patients (median age 67.9 years, range 59 to 78) required correction of misplaced or dislocated endografts in the thoracic (n = 1) or infrarenal abdominal aorta (n = 6). In 1 patient, an infrarenal bifurcated stent-graft was mistakenly deployed across a renal artery; repositioning was accomplished by tugging caudally on a guidewire placed across the endograft bifurcation and exteriorized from both femoral arteries. An inflated balloon catheter was used to reposition 3 dislocated aortic devices (1 thoracic, 2 infrarenal) and 3 iliac graft limbs that had disconnected from the main graft body 6 to 12 months after implantation. RESULTS: Repositioning maneuvers were successful in all cases, with the devices being moved from 5 to 27 mm (median 7.8 mm). There were no procedure-related complications. CONCLUSIONS: Nonsurgical repositioning of misplaced aortic prostheses is technically feasible in individual cases. The risk associated with the procedure, however, cannot yet be evaluated.
No takes yet. Share an insight, caveat, or question.
Görich et al. (2000) conducted a case report in Malpositioned or dislocated aortic endografts (n=7). Percutaneous pull-down maneuvers was evaluated on Successful repositioning of the device. Percutaneous pull-down maneuvers successfully repositioned malpositioned or dislocated aortic endografts in all 7 cases (100%) with no procedure-related complications.