Key result
Patient radiation exposure during EVAR was nearly 50% higher when performed in an angiosuite compared to an operating theatre.
Why the study?
Does performing EVAR in an angiosuite compared to an operating theatre increase patient radiation exposure?
Observational (n=146)
Does performing EVAR in an angiosuite compared to an operating theatre increase patient radiation exposure?
Performing EVAR in an angiosuite results in nearly 50% higher patient radiation exposure compared to an operating theatre, a risk that must be weighed against the technical advantages of the angiosuite.
Answers are grounded in this paper and cite their sources.
Higher angiosuite radiation during EVAR warrants caution; leaves open optimal setting pending randomized data.
The aim of this study was to assess the patient dosimetric impact of endovascular abdominal aortic aneurysm repair (EVAR), both in an operating theatre (OR) and in an angiosuite (AS), with the facility of Dynamic CT (Dyna CT, Siemens AG, Berlin, Germany). One hundred and forty-six consecutive EVAR procedures dating from May 2011 to March 2013 were analysed. These were performed either in an OR (n = 97) using a mobile C-arm or in an AS (n = 49) equipped with a ceiling-mounted angiography system. Air kerma area product (P(KA)) and total air kerma at reference point (K(a,r)) values were reported for all procedures and Dyna CT. Radiation exposure during EVAR was quite low in the majority of patients but nearly 50 % higher if performed in AS vs. OR. Median Dyna CT K(a,r) was the same as an entire EVAR procedure in OR. The higher patient's radiation exposure recorded in the AS should be balanced with the technical advantages given to the EVAR procedure.
No takes yet. Share an insight, caveat, or question.
Bruschi et al. (2014) conducted an observational in Abdominal aortic aneurysm (n=146). EVAR in an angiosuite (AS) vs. EVAR in an operating theatre (OR) was evaluated on Patient radiation exposure (Air kerma area product and total air kerma at reference point) (nearly 50% higher). Patient radiation exposure during EVAR was nearly 50% higher when performed in an angiosuite compared to an operating theatre.
Synapse has enriched 10 related papers. Consider them for comparative context: