Key result
Implantation of the INTACT bioprosthesis in 167 patients resulted in 5% hospital mortality, with a 22% 3-year actuarial rate of calcification in patients under 20 and no valve failures in adults.
Why the study?
Does the INTACT bioprosthesis improve durability and reduce calcification in patients undergoing valve replacement?
Observational (n=167)
Does the INTACT bioprosthesis improve durability and reduce calcification in patients undergoing valve replacement?
Answers are grounded in this paper and cite their sources.
The INTACT bioprosthesis showed satisfactory early hemodynamics and no failures in adults, but a 22% 3-year calcification rate in patients under 20 and a risk of ventricular perforation due to its high-profile stent.
The INTACT (previously Xenotech) valve incorporates three major departures from previously available bioprostheses. The following innovations are aimed at reducing calcification and improving durability: (1) the stent has a high profile configuration and the sinuses are deeply excavated. This should maintain normal coaption of the leaflets and prevent cusp abrasion; (2) stress-free fixation preserves the structural characteristics of the leaflets; (3) the valve is treated with toluidine blue, which occupies the calcium binding sites. During a 36-month period 185 valves were implanted in 167 patients. The procedure used most often was isolated mitral valve replacement. Twenty-six patients were under the age of 20. There were 8 hospital deaths (5% mortality) and 12 late deaths. Two of the hospital deaths were the result of ventricular perforation, and it is believed that the high profile of the stent contributed. In the group aged less than 20 years, there have been three episodes of valve calcification giving a 3-year actuarial rate for calcification of 22%. In the adult group, there have been no valve failures, and postoperative catheterization studies in this group demonstrated satisfactory hemodynamics.
No takes yet. Share an insight, caveat, or question.
M.A. Williams (1988) conducted an observational in Valvular heart disease requiring replacement (n=167). INTACT bioprosthesis was evaluated on Hospital mortality. Implantation of the INTACT bioprosthesis in 167 patients resulted in 5% hospital mortality, with a 22% 3-year actuarial rate of calcification in patients under 20 and no valve failures in adults.
Synapse has enriched 10 related papers. Consider them for comparative context: