A hydromechanical simulator was successfully developed to evaluate the dynamic actions of mitral valve prostheses, enabling detailed analysis of valve open areas, pressures, and flows.
Answers are grounded in this paper and cite their sources.
Enables detailed in vitro mitral prosthesis testing; leaves open whether simulator data predict clinical performance.
A hydromechanical simulator of the human left heart and systemic circulation system has been built for evaluating the dynamic actions of mitral valve prostheses. The facility includes a transparent and compliant model ventricle pumping into a simulation of the systemic circulation impedance. Transparent materials permit detailed high speed photography of valve action from the left ventricular aspect. This action has been analyzed by digitizing the valve open area and computing the variation of area with time. Additionally, pressures across and flows through the valves have been measured. Comparison has been made between the maximum observed valve open areas and the areas computed using the appropriate pressure and flow measurements substituted in clinically used equations.
No takes yet. Share an insight, caveat, or question.
Scotten et al. (1979) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: