Key result
A novel sensorless rotational speed-based control algorithm for CFLVAD provided adequate perfusion while preventing ventricular suction across various simulated conditions, including up to 6% noise.
Why the study?
The authors sought to develop a novel continuous flow left ventricular assist device control algorithm to maintain physiologic perfusion while avoiding ventricular suction using only pump speed without model-based estimation.
Does a sensorless rotational speed-based control algorithm maintain physiologic perfusion and prevent ventricular suction in an in silico model of a CFLVAD compared to other control strategies?
Does a sensorless rotational speed-based control algorithm maintain physiologic perfusion and prevent ventricular suction in an in silico model of a CFLVAD compared to other control strategies?
A novel sensorless rotational speed-based control algorithm for CFLVADs successfully maintains physiologic perfusion and prevents suction in silico without requiring direct flow or pressure measurements.
May support suction-free CFLVAD control using pump speed alone; leaves open need for randomized validation before clinical use.
OBJECTIVE: Continuous Flow Left Ventricular Assist Devices (CFLVAD) are circulatory support devices that are implanted in patients with end-stage heart failure. We developed a novel control algorithm for CFLVAD to maintain physiologic perfusion while avoiding ventricular suction using only the intrinsic pump measurement of pump speed and without utilizing model-based estimation. METHODS: The controller objective is to maintain a differential pump speed setpoint. A mathematical model of the circulatory system coupled with a model of a CFLVAD was used to test the control algorithm in silico. Robustness and efficacy were evaluated by comparing the proposed control algorithm to constant speed control, differential pump pressure control, mean aortic pressure control, and ventricular end diastolic pressure control during (1) rest and exercise conditions, (2) a rapid eight-fold increase in pulmonary vascular resistance under rest and exercise, (3) transitions from rest to exercise, and exercise to rest, (4) safe mode during left ventricular asystole, and (5) RPM measurement noise of 1% to 10% for (1) to (4). RESULTS AND CONCLUSION: The control algorithm provided adequate perfusion while preventing ventricular suction for all test conditions. Performance did not deteriorate significantly with pump speed measurement noise of up to 6%. The safe mode successfully detected asystole and maintained adequate perfusion to sustain life even when the differential pump speed was low. SIGNIFICANCE: Maintaining a constant differential pump speed can simultaneously achieve physiologic perfusion and suction prevention without needing unreliable, direct measurements of flow or pressure, or complex parameter or model-based estimation techniques.
No takes yet. Share an insight, caveat, or question.
Meki et al. (2019) studied End-stage heart failure. Sensorless rotational speed-based control algorithm for CFLVAD vs. Constant speed control, differential pump pressure control, mean aortic pressure control, and ventricular end diastolic pressure control was evaluated on Maintenance of physiologic perfusion while avoiding ventricular suction. A novel sensorless rotational speed-based control algorithm for CFLVAD provided adequate perfusion while preventing ventricular suction across various simulated conditions, including up to 6% noise.