Key result
A portable INR/PT monitor was accurate compared with clinical laboratory determinations, with approximately 85% of measurements within 0.9 INR units of the corresponding laboratory value.
Why the study?
Does a portable INR/PT monitor provide accurate measurements compared to clinical laboratory determinations in outpatients?
Does a portable INR/PT monitor provide accurate measurements compared to clinical laboratory determinations in outpatients?
Portable INR/PT monitors provide accurate measurements comparable to clinical laboratories, facilitating rapid warfarin dose adjustments in outpatient settings.
Answers are grounded in this paper and cite their sources.
May support portable INR monitoring in outpatients; leaves open prospective validation for clinical outcomes.
We review our collective experience over the past 4 years using a portable instrument that measures the International Normalized Ratio (INR) and the prothrombin time (PT). The device can be used readily in ambulatory clinic settings and by patients in their homes. The major advantage of the monitor is that the INR/PT can be measured quickly; if a change in the dose of warfarin is necessary, instructions can be discussed directly with the patient at the time of the visit. Compared with clinical laboratory determinations, monitor INRs are quite accurate, with approximately 85% of measurements within 0.9 INR units of the corresponding laboratory value. If necessary, monitor INRs can be adjusted to approximate a specific clinical laboratory's INRs with a regression formula derived from analysis of duplicate INRs using the monitor and the laboratory. Although the Food and Drug Administration has not yet approved use of the monitor by patients at home, use of a portable measuring device in the home by home health agencies may be valuable in selected cases.
No takes yet. Share an insight, caveat, or question.
White et al. (1994) conducted a review in Patients requiring INR/PT monitoring. Portable INR/PT monitor vs. Clinical laboratory determinations was evaluated on Measurements within 0.9 INR units of the corresponding laboratory value. A portable INR/PT monitor was accurate compared with clinical laboratory determinations, with approximately 85% of measurements within 0.9 INR units of the corresponding laboratory value.
Synapse has enriched 10 related papers. Consider them for comparative context: