The benchtop blood analyzer is the gold standard for blood oxygen saturation (SO 2 ) and hemoglobin (Hb) analysis. However, the benchtop analyzer only provides values at a given point in time. In the field of cardiovascular perfusion and the practice of cardiopulmonary bypass (CPB), continuous measurement of SvO 2 and hemoglobin values have become commonplace. Two devices currently available which monitor these values are the Terumo CDI 500 and Spectrum Medical M4. A retrospective study was conducted to examine the accuracy of the M4 technology and the CDI 500 as they compare to each other and the ABL90 FLEX, a benchtop blood gas analyzer. The data revealed the magnitude of mean differences were small, even when significant. However, the 95% Limits of Agreement were too large for either device to allow substitution of the CDI 500 and M4 hemoglobin or SvO 2 values for ABL90 values. As recommended by the manufacturers, the CDI 500 and M4 should only be used as a trending device.