Key result
The application of unipolar electrocautery during a pacemaker pulse generator change caused a sudden and unexpected loss of ventricular capture lasting 30 seconds.
Why the study?
Although electrocautery is generally considered safe during cardiac device procedures, device makers have variable recommendations and it is rarely known to cause device failure.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
Unipolar electrocautery during pacemaker generator replacement can cause transient, unexpected loss of ventricular capture, emphasizing the need for backup pacing and closed-loop communication in pacemaker-dependent patients.
In the advent of increasing benefits of cardiac devices, more and more implants are being done. Pacing devices reaching the end of service need to be changed. The use of electrocautery (EC) to maintain hemostasis during cardiac device implantation is efficient and safe. Device makers have variable recommendations for the use of EC. Generally, considered safe, EC has been rarely known to cause device failure. We describe a case of a dual-chamber device, pulse generator change, where EC caused a sudden, unexpected loss of pacing function that lasted for 30 seconds. This case report highlights the gaps in the process of undertaking these high-risk changes.
No takes yet. Share an insight, caveat, or question.
Qureshi et al. (2022) conducted a case report in Complete atrioventricular block (n=1). Unipolar electrocautery was evaluated on Loss of ventricular capture. The application of unipolar electrocautery during a pacemaker pulse generator change caused a sudden and unexpected loss of ventricular capture lasting 30 seconds.
Synapse has enriched 10 related papers. Consider them for comparative context: