Why the study?
Does unipolar surgical electrocautery cause runaway pacemaker syndrome in patients with single chamber ventricular pacemakers?
Does unipolar surgical electrocautery cause runaway pacemaker syndrome in patients with single chamber ventricular pacemakers?
Unipolar surgical electrocautery can induce runaway pacemaker syndrome, necessitating emergent explantation of the pulse generator.
Answers are grounded in this paper and cite their sources.
Warrants caution with unipolar electrocautery near pacemakers; leaves open generalizability to modern devices or bipolar alternatives.
A case of a unipolar surgical electrocautery-induced runaway pacemaker is described. The single chamber ventricular pacemaker began firing at 140 beats/min with intermittent capture and the abnormal rhythm persisted after cessation of the electrocautery. Emergent explanation of the pulse generator was required.
No takes yet. Share an insight, caveat, or question.
Louis I. Heller (1990) studied this question.