Key result
Monopolar electrocautery triggers VF during temporary AV sequential pacing in Ebstein's anomaly.
Why the study?
Arrhythmias and pacemaker malfunction due to monopolar electrocautery use in cardiac surgical patients with Ebstein’s anomaly on temporary pacemakers are clinically significant but not well documented.
Case Report (n=1)
Monopolar electrocautery can induce ventricular fibrillation in patients on temporary pacemakers, suggesting ultrasonic harmonic cautery may be a safer alternative.
Answers are grounded in this paper and cite their sources.
Alerts to VF risk with monopolar electrocautery in temporary-paced Ebstein's patients; leaves open whether harmonic scalpel is safer in broader populations.
Electrocautery Induced Ventricular Fibrillation in a Cardiac Surgical Patient on Temporary Pacemaker: Case Report Arrhythmias occur commonly following Cardiopulmonary bypass. Patients with Ebstein’s anomaly commonly develop Atrio-ventricular nodal block post-operatively due to injury to the sinus node. Atrioventricular sequential pacemaker is indicated in this situation. Arrhythmias and pacemaker malfunction can occur due to the use of monopolar electrocautery. Here we are presenting a case of Ebstein’s anomaly undergoing corrective surgery developed ventricular fibrillation due to the use of electrocautery when the patient was on Atrio-ventricular sequential pacing. We revived the patient successfully and use of ultrasonic harmonic cautery is safe in these cases.
No takes yet. Share an insight, caveat, or question.
Mridu Paban Nath (2013) conducted a case report in Ebstein's anomaly (n=1). Monopolar electrocautery was evaluated on Ventricular fibrillation. Monopolar electrocautery induced ventricular fibrillation in a patient with Ebstein's anomaly on temporary atrio-ventricular sequential pacing.