Runaway pacemaker malfunction, though rare, can still occur in modern CPU-controlled pacemakers and cause potentially life-threatening dysrhythmia.
Answers are grounded in this paper and cite their sources.
Clinicians should remain vigilant for runaway pacemaker in symptomatic patients with modern devices; leaves open questions on malfunction frequency and surveillance.
"Runaway" pacemaker is a rare entity that occurs when a malfunctioning artificial cardiac pacemaker abruptly accelerates its pacing rate above the set upper rate limit. This can result in life-threatening dysrhythmia. Runaway pacemaker used to occur more frequently in older model pacemakers, but now with newer pacemaker generators, runaway pacemaker is a rare entity. We report the case of a runaway pacemaker in a modern CPU-controlled pacemaker in a 79-year-old man presenting with lightheadedness and review the literature regarding this rare entity.
No takes yet. Share an insight, caveat, or question.
Makaryus et al. (2005) studied this question.