Highlights the risk of runaway pacemaker syndrome and intermittent nonoutput failure as manifestations of battery end-of-life in VVI Intermedics pacemakers, warranting close follow-up.
Answers are grounded in this paper and cite their sources.
Alerts clinicians to monitor aging Intermedics VVI devices for runaway behavior; single case leaves generalizability open.
A case of a runaway pacemaker with intermittent nonoutput failure of a VVI Intermedics pacemaker is presented. This is the first report of such a phenomenon in that type of pacemaker with no apparent reason other than approaching battery end-of-life. The reported case warrants a close follow-up of patients with the same type of device.