Runaway pacemaker is a rare but potentially lethal complication requiring immediate pacemaker exchange and emergency maneuvers like magnet placement.
Answers are grounded in this paper and cite their sources.
Runaway pacemaker may trigger fatal VT requiring urgent magnet use; case reports leave open standardized protocols for validation.
Runaway pacemaker is a rare, but still existing potential lethal complication in permanent pacemakers. Within 4 1/2 years, we saw two cases of runaway pacemaker in patients with multiprogrammable, VVI pacemakers (Siemens-Elema, Model 668). In both cases a pacemaker-induced ventricular tachycardia (rate 240-260 beats/min) was documented. One patient died. Runaway pacemakers must be exchanged as soon as possible. Until this can be accomplished, different emergency maneuvers should be tried. As documented in the cases presented, placing a magnet over the pacemaker may result in a lower, more physiological pacing rate. Reprogramming the pulse generator to a lower output or the use of external chest wall overdrive stimulation may also be successful, but these procedures require the presence of an adequate escape rhythm. If this is not the case or the former maneuvers have failed, an external pacemaker may be connected to the permanent pacing lead. Thereafter, the lead can be safely cut. As an alternative, a temporary transvenous pacing lead may be established prior to disconnecting the permanent pacing lead.
No takes yet. Share an insight, caveat, or question.
Mickley et al. (1989) studied this question.