Key result
Emergency pulse generator replacement resolves runaway pacemaker stimuli caused by undetected battery depletion.
Why the study?
Runaway pacemaker malfunction due to undetected battery depletion is an uncommon and potentially lethal circuit malfunction requiring clinical awareness.
Case Report (n=2)
Undetected battery depletion in single chamber pacemakers can lead to potentially lethal runaway pacemaker malfunction, necessitating emergency pulse generator replacement.
Answers are grounded in this paper and cite their sources.
May prompt device interrogation for episodic dizziness in patients with older Medtronic pacemakers; leaves open the incidence of runaway malfunction.
Runaway pacemaker is an uncommon, potentially lethal circuit malfunction characterized by sudden onset of erratic pacing at rapid nonphysiological rates. Two patients with a single chamber pacemaker (Medtronic ST 8331 and 8419) presented with episodic dizziness. ECG revealed recurrent decrescendo amplitude episodes of runaway stimuli at 2,400 and 2,600 ppm, approximately 3 seconds in duration, separated by pacing at 62.5 and 65 ppm, respectively. Fortunately the runaway stimuli were subthreshold and did not result in capture of the ventricle. Emergency pulse generator replacement was uneventful. Both leads were normal and both pulse generators had low battery voltages at 1.488 and 1.78 V, respectively.
No takes yet. Share an insight, caveat, or question.
Vijayaraman et al. (2002) conducted a case report in Runaway pacemaker malfunction (n=2). Undetected battery depletion was evaluated. Undetected battery depletion in two patients with single chamber pacemakers caused runaway stimuli at 2,400 and 2,600 ppm, which resolved uneventfully after emergency pulse generator replacement.