At a phantom peak velocity of 0.03 cm/s, the error in Tissue Velocity Imaging estimation ranged from 31% to 313% depending on the pulse repetition frequency setting.
Why the study?
Does pulse repetition frequency setting affect the accuracy of Tissue Velocity Imaging in estimating very low tissue velocities?
Does pulse repetition frequency setting affect the accuracy of Tissue Velocity Imaging in estimating very low tissue velocities?
The PRF setting is crucial for the accuracy of Tissue Velocity Imaging at very low tissue velocities (0.03 cm/s) and should be reduced when analyzing low-level muscle contractions.
Answers are grounded in this paper and cite their sources.
TVI errors at 0.03 cm/s velocities vary widely with PRF; hypothesis-generating for protocol optimization in low-velocity tissue analysis.
BACKGROUND: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal tissue velocities can be very low. In particular, when performing isometric contractions or contractions of low force level the velocities may be much lower compared to the myocardial tissue velocities. METHODS: In this study TVI was evaluated for estimation of tissue velocities below the typical myocardial velocities. An in-house phantom was used to see how different PRF-settings affected the accuracy of the velocity estimations. RESULTS: With phantom peak velocity at 0.03 cm/s the error ranged from 31% up to 313% with the different PRF-settings in this study. For the peak velocities at 0.17 cm/s and 0.26 cm/s there was no difference in error with tested PFR settings, it is kept approximately around 20%. CONCLUSIONS: The results from the present study showed that the PRF setting did not seem to affect the accuracy of the velocity estimation at tissue velocities above 0.17 cm/s. However at lower velocities (0.03 cm/s) the setting was crucial for the accuracy. The PRF should therefore preferable be reduced when the method is applied in low-level muscle contraction.
No takes yet. Share an insight, caveat, or question.
Lindberg et al. (2013) studied skeletal muscle low-level contractions (simulated). Tissue Velocity Imaging (TVI) with different PRF settings vs. True phantom velocity was evaluated on Accuracy of velocity estimation (mean error %). At a phantom peak velocity of 0.03 cm/s, the error in Tissue Velocity Imaging estimation ranged from 31% to 313% depending on the pulse repetition frequency setting.
Synapse has enriched 10 related papers. Consider them for comparative context: