September 1, 2002European Heart Journal - Cardiovascular ImagingOpen Access

Analysis of Temporal Requirements for Myocardial Tissue Velocity Imaging

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does reducing the sampling rate of tissue velocity imaging affect the accuracy of myocardial velocity measurements in healthy individuals?

Population

16 healthy individuals

Comparison

Off-line reduction of tissue velocity imaging… vs Initial high frame rate (141-203 frames/s)

Design

Other

Authors

BLBritta LindKarolinska University HospitalJNJacek NowakImmungenetics (Germany)JDJ. DorphKarolinska University Hospital

Discussion

Loading...

Member takes

Overview

High frame rates are needed to avoid underestimating myocardial velocities; hypothesis-generating for standardized protocols in clinical echocardiography.

Structured PICO

Does reducing the sampling rate of tissue velocity imaging affect the accuracy of myocardial velocity measurements in healthy individuals?

P
Population
16 healthy individuals
I
Intervention
Off-line reduction of tissue velocity imaging sampling rate to 50%, 25%, and 12.5% of the initial frame rate
C
Comparator
Initial high frame rate (141-203 frames/s)
O
Outcome
Deviation of myocardial velocity profile measurements (peak systolic and diastolic velocities) from initial valuessurrogate

A minimum frame rate of 70-100 frames/s is required for accurate tissue velocity imaging measurements to prevent significant deviation and underestimation of peak velocities.

Cite This Study

Lind et al. (2002) studied this question.

synapsesocial.com/papers/6a74d7fa54045e727a09101chttps://doi.org/10.1053/euje.3.3.214
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Regional diastolic function in ischaemic heart disease using pulsed wave Doppler tissue imaging1999 · 202 citations
  2. 2Assessment of regional long-axis function during dobutamine echocardiography2001 · 22 citations
  3. 3Reliability of Tissue Doppler Wall Motion Monitoring After Heart Transplantation for Replacement of Invasive Routine Screenings by Optimally Timed Cardiac Biopsies and Catheterizations2001 · 162 citations
  4. 4Communication In The Presence Of Noise1998 · 4,873 citations