Does electronic IVUS compare to mechanical IVUS in accurately quantifying mild to moderate atherosclerosis in human femoral arteries?
Does electronic IVUS compare to mechanical IVUS in accurately quantifying mild to moderate atherosclerosis in human femoral arteries?
Answers are grounded in this paper and cite their sources.
Both electronic and mechanical IVUS devices show equivalent performance but lack sufficient accuracy for determining mild to moderate intimal thickening in peripheral arteries.
To define the accuracy of electronic and mechanical ultrasound (US) devices for determining the thickness of intima and media, 32 fresh normal and atherosclerotic human femoral arteries were obtained at necropsy. The samples were imaged with a 64-element array and a mechanically rotating US transducer at 20 MHz. The mean thickness of the intimal and medial layer was measured with electronic calipers followed by histopathologic and micromorphometric analysis. Morphometric correlation for intima showed r = 0.64 for the electronic and r = 0.58 for the mechanical US device. The correlation between ultrasonic and histologic measurement of medial thickness in normal and diseased specimens was r = 0.79 for the multielement and r = 0.76 for the mechanical transducer. In conclusion, multielement array transducers are equivalent to mechanically driven probes in the quantitative assessment of peripheral arterial wall layers. Both systems lack sufficient accuracy in the determination of mild to moderate intimal thickening.
No takes yet. Share an insight, caveat, or question.
Duda et al. (1994) studied this question.