Why the study?
Does intravascular ultrasound imaging accurately measure arterial wall thickness compared to histology in human atherosclerotic arteries?
Does intravascular ultrasound imaging accurately measure arterial wall thickness compared to histology in human atherosclerotic arteries?
Intravascular ultrasound with a 20 MHz transducer provides accurate measurements of microanatomic arterial dimensions and atheromatous involvement compared to histology.
Answers are grounded in this paper and cite their sources.
Supports IVUS validation in animal models; leaves open translation to human atherosclerotic arteries.
A prototype intravascular ultrasound imaging catheter with a 20 MHz transducer was used to obtain 59 cross-sectional images in 14 segments of human atherosclerotic arteries. Three distinct components of the arterial wall were visualized on the ultrasound images: a highly reflective intima, an echolucent media, and a moderately reflective adventitia. Images were obtained at 1 mm increments in vitro and were compared with histologic sections at the same levels. Measurements of the arterial layers showed a close correlation between ultrasound images and histologic sections for the thickness of the intimal plaque (r = 0.91), the media (r = 0.83), and the total wall thickness (r = 0.85). The ultrasound images overestimated the mean intimal and total wall thickness by 0.3 mm and 0.7 mm compared to measurements in histologic sections (p less than 0.001). Intravascular imaging with high-frequency ultrasound is an accurate method for measuring microanatomic arterial dimensions and the extent of atheromatous involvement of the arterial wall. This method could represent an important adjunct to traditional angiographic techniques for assessing the severity of atherosclerosis.
No takes yet. Share an insight, caveat, or question.
Mallery et al. (1990) studied this question.