Key result
Spot electrodes and band electrodes showed no differences in determining stroke volume and systolic time interval measures across different tasks, with correlations around 0.8.
Why the study?
Do spot electrodes differ from band electrodes in determining systolic time intervals and stroke volume during impedance cardiography?
Cross-Sectional (n=12)
Do spot electrodes differ from band electrodes in determining systolic time intervals and stroke volume during impedance cardiography?
Spot electrodes perform similarly to band electrodes for measuring stroke volume and systolic time intervals via impedance cardiography across various tasks.
Answers are grounded in this paper and cite their sources.
Spot electrodes may simplify impedance cardiography setups; leaves open larger validation trials before routine clinical use.
Impedance cardiography has become of widespread interest as a noninvasive cardiovascular monitoring technique. This study compared the use of spot electrodes and two different types of band electrodes in the determination of systolic time intervals and stroke volume. EKG, impedance cardiogram, and phonocardiogram were recorded on 12 volunteers during rest, during a reaction time task, and after a short exercise task. Systolic time intervals (pre-ejection period and left ventricular ejection time) were computed on a beat-to-beat basis and on the ensemble-averaged signals. The only differences between spot and band electrodes were for impedance base level and electrode conductance. For stroke volume and the systolic time interval measures there were no differences between electrode arrays, nor were there any interactions with tasks. Correlations for stroke volume and systolic time interval measures as determined under different electrode conditions were around .8.
No takes yet. Share an insight, caveat, or question.
Boomsma et al. (1989) reported a cross-sectional. Spot electrodes vs. Band electrodes was evaluated on Systolic time intervals and stroke volume. Spot electrodes and band electrodes showed no differences in determining stroke volume and systolic time interval measures across different tasks, with correlations around 0.8.