Key result
Testing plasma specimens on four different troponin I assays resulted in 44% of patients being categorized differently by instrument-specific cutoff values, despite high correlations (R2≥0.94).
Why the study?
Does the choice of troponin I assay platform and cutoff value affect patient categorization in suspected acute myocardial infarction?
Observational
Does the choice of troponin I assay platform and cutoff value affect patient categorization in suspected acute myocardial infarction?
Different troponin I assays and cutoff values lead to significant reclassification of patients, highlighting the need for clinician education when changing laboratory platforms.
Supports caution with assay platform switches in suspected MI; leaves open standardization effects on diagnostic accuracy.
Measurement of troponin concentration is important for evaluating patients with suspected acute myocardial infarction or cardiac damage, with results interpreted relative to method-specific cutoff values. This study evaluates changes in sample categorization and patient management based on different test methods. Plasma specimens were tested for troponin I on 4 different instrument platforms, with chart review. Results were compared with cutoff values corresponding to the 99th percentile reference interval and the concentration having 10% coefficient of variation. Significant correlations were observed among instrument platforms (R2≥0.94). However, 44% of patients were categorized differently by instrument-specific cutoff values. The use of higher sensitivity assays was associated with 47% to 106% increase in number of results exceeding both the 99th percentile and 10% CV cutoffs. Changing instrument platforms can cause reclassifications of patient results. Laboratories adopting new troponin measurement methods should perform clinician education regarding its interpretation and the importance of serial testing.
No takes yet. Share an insight, caveat, or question.
Bonham et al. (2009) conducted an observational in Suspected acute myocardial infarction or cardiac damage. Four commercially available troponin I assays vs. Different instrument platforms was evaluated on Sample categorization based on different test methods. Testing plasma specimens on four different troponin I assays resulted in 44% of patients being categorized differently by instrument-specific cutoff values, despite high correlations (R2≥0.94).