Key result
A next-generation α-Gal-free tissue valve demonstrated good haemodynamic status and function without immune reaction or calcification at 18 months in a sheep model of mitral valve replacement.
Why the study?
Does a novel α-Gal-free tissue valve with multiple anticalcification treatments demonstrate good function and reduced calcification in a sheep model of mitral valve replacement?
Does a novel α-Gal-free tissue valve with multiple anticalcification treatments demonstrate good function and reduced calcification in a sheep model of mitral valve replacement?
Answers are grounded in this paper and cite their sources.
A novel α-Gal-free tissue valve utilizing multiple anticalcification and immunological modification therapies showed promising durability and absence of calcification in a preclinical sheep model.
OBJECTIVES: Conventional crosslinking with glutaraldehyde (GA) renders cardiac xenografts inert, non-biodegradable and non-antigenic, but is a main cause for dystrophic calcification due to phospholipids, free aldehyde groups and residual antigenicity. A significant immune reaction to the galactose-α-1,3 galactose β-1,4-N-acetylglucosamine (α-Gal) of a GA-fixed cardiac xenograft occurs, leading to calcification. We developed a next-generation α-Gal-free tissue valve with GA-fixed cardiac xenografts, treated using a novel combined anticalcification protocol including immunological modification, which was demonstrated effective in a small animal study. METHODS: Porcine aortic valves were decellularized with 1% sodium dodecyl sulphate, 1% Triton X-100 and 1% sodium lauroyl sarcosinate and immunologically modified with α-galactosidase. The valves were treated by a polyethylene glycol space filler, fixed with GA in 75% ethanol + 5% octanol and detoxified with glycine. We manufactured the tissue valve with the porcine aortic valve mounted on a Nitinol (nickel-titanium memory alloy) plate. The tissue valve was placed under in vitro mock circulation, and durability from mechanical stress was evaluated for 100 days. Ten sheep underwent mitral valve replacement with the tissue valve, and haemodynamic, radiological, immunohistopathological and biochemical results were obtained for 18 months after implantation. RESULTS: The in vitro circulation experiment demonstrated that the valve functioned well with good morphology. Eight sheep survived for 1, 2, 5, 10, 14, 53, 546 and 552 days after mitral valve replacement, but two sheep did not survive. An evaluation by echocardiography and cardiac catheterization demonstrated good haemodynamic status and function of the mitral valve at 18 months after implantation. The xenografts were well preserved without a α-Gal immune reaction or calcification based on the immunological, radiographic, microscopic and biochemical examinations. CONCLUSIONS: We developed a next-generation α-Gal-free tissue valve with simultaneous use of multiple anticalcification therapies and novel tissue treatments such as decellularization, immunological modification with α-galactosidase, space filler, an organic solvent and detoxification. Future investigations should evaluate α-Gal-free substitutes such as our tissue valve, and a future clinical study is warranted based on these promising preclinical results.
No takes yet. Share an insight, caveat, or question.
Lim et al. (2014) studied Mitral valve replacement (animal model) (n=10). Next-generation α-Gal-free tissue valve was evaluated on Haemodynamic, radiological, immunohistopathological and biochemical results. A next-generation α-Gal-free tissue valve demonstrated good haemodynamic status and function without immune reaction or calcification at 18 months in a sheep model of mitral valve replacement.