Key result
A 71-year-old woman experienced an asymptomatic anterior perforation of an ICD lead into subcutaneous tissues one month post-implantation, requiring surgical removal.
Case Report (n=1)
This case highlights the possibility of delayed, asymptomatic ICD lead perforation presenting with loss of pacing and sensing, requiring imaging for confirmation and surgical extraction.
Answers are grounded in this paper and cite their sources.
Raises awareness for delayed asymptomatic ICD lead perforation; single case leaves open incidence, risk factors, and surveillance needs.
A 71-year-old woman underwent routine implantable cardioverter defibrillator implantation. On a predischarge check the next day, electrical signals and thresholds were excellent and similar to those at implant. The chest X-ray was unremarkable and showed good lead position at the right ventricular apex (RVA). At a routine one-month postimplant visit, electrograms were found to be miniscule, and pacing could not be achieved. Chest X-ray and fluoroscopy suggested perforation, then this was confirmed by computed tomography scan. The tip of the lead was estimated to be within 7 mm of the surface of the skin. The system was removed surgically, and the patient continued to do well.
No takes yet. Share an insight, caveat, or question.
Fisher et al. (2007) conducted a case report in ICD lead perforation (n=1). Implantable cardioverter defibrillator implantation was evaluated on Asymptomatic anterior perforation of an ICD lead. A 71-year-old woman experienced an asymptomatic anterior perforation of an ICD lead into subcutaneous tissues one month post-implantation, requiring surgical removal.