Key result
Delayed, asymptomatic RV perforation by an ICD lead resolves with successful repositioning.
Why the study?
Delayed perforation of ICD leads can be subclinical but may cause major clinical manifestations, and strategies to prevent and detect it are needed.
Case Report (n=1)
Highlights the possibility of delayed, asymptomatic ICD lead perforation and the potential utility of telemonitoring and septal placement to prevent complications.
Subacute ICD lead perforation warrants prompt CT confirmation; extends Level 5 case documentation of migration risks.
The figure shows a single-chamber defibrillator [implantable cardioverter-defibrillator (ICD)] with a dual-coil, passive-fixation lead (Riata 1570, St Jude Medical, MN, USA) perforating the right ventricular apex with migration of the electrode-tip to the costophrenic angle, as suggested by the chest X-ray ( Panel A ) and confirmed by the computed tomographic (CT) scan ( Panels B and C ). This event occurred subacutely in a 70-year-old woman implanted in secondary prevention for non-ischaemic dilated cardiomyopathy. It was diagnosed 3 months after ICD implantation (lack of ventricular sensing/capture) without the presence of any symptom or ICD-related finding (recording of false ventricular tachycardia and/or diaphragmatic stimulation). Lead repositioning was performed in the electrophysiology lab ( Panel D ) without acute/late complications. Delayed perforation is probably not so rare if we consider subclinical events reported in CT studies without modification of pacing/ICD systems performance, but it can be associated with major clinical manifestations: inappropriate shock, abdominal pain, diaphragm stimulation, chest pain, syncope, and tamponade. However, considering the potential risk of similar events, strategies targeted at preventing perforation (e.g. placement of the lead towards interventricular septum, especially in patients at higher risk of perforation) and avoiding delayed diagnosis, like adopting telemonitoring systems, could help in reducing possible complications. Conflict of interest: none declared.
No takes yet. Share an insight, caveat, or question.
Diemberger et al. (2010) conducted a case report in Non-ischaemic dilated cardiomyopathy with ICD lead perforation (n=1). Implantable cardioverter-defibrillator (ICD) implantation was evaluated on Delayed asymptomatic migration of ICD lead. A 70-year-old woman experienced delayed, asymptomatic perforation of the right ventricular apex by an ICD lead 3 months after implantation, which was successfully repositioned without complications.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: