Key result
Methamphetamine cardiomyopathy is linked to ~9-J higher defibrillation thresholds during ICD implantation.
Why the study?
Does methamphetamine cardiomyopathy increase defibrillation thresholds and the requirement for high-output devices in patients undergoing ICD implantation compared to matched controls?
Case-Control (n=20)
No
Does methamphetamine cardiomyopathy increase defibrillation thresholds and the requirement for high-output devices in patients undergoing ICD implantation compared to matched controls?
Answers are grounded in this paper and cite their sources.
Patients with methamphetamine cardiomyopathy undergoing ICD implantation have significantly elevated defibrillation thresholds and frequently require high-output devices, warranting intraoperative DFT testing and consideration of empiric high-energy ICD generators.
INTRODUCTION: Identification of patients with an increased risk of high defibrillation thresholds (DFTs) is important in planning implantable cardioverter-defibrillator (ICD) procedures. Clinical observations have suggested that patients with methamphetamine cardiomyopathy (MACMP) have significantly elevated defibrillation thresholds. We hypothesized that MACMP patients would have higher DFT thresholds than controls and would require procedural changes during ICD implantation to accommodate higher thresholds. METHODS: We identified consecutive patients with MACMP undergoing ICD implantation at the academic center from 2003 to 2007. We then compared DFTs against age-and sex-matched controls. RESULTS: The MACMP (n = 10) group showed significantly increased DFT thresholds (23.7 ± 6.7 J) compared with age and sex-matched controls (14.5 ± 4.6 J, p < 0.005). Additionally, patients with MACMP had evidence of more severe congestive heart failure, with increased B-type natrieutic protein (BNP) levels (1173 ± 784 vs 260 ± 349, p = 0.02) and decreased left ventricular ejection fraction (LVEF) (17.8 ± 9.4 vs 35.9 ± 15.2, p = 0.02). MACMP patients required high output devices than controls (50% versus 0%, p = 0.03). Differences between groups remained significant despite adjusting for LVEF. CONCLUSIONS: Planning for ICD implantation should take into consideration a history of methamphetamine abuse, mandating DFT testing and empiric consideration of high output devices for such patients.
No takes yet. Share an insight, caveat, or question.
Malhotra et al. (2017) conducted a case-control in Cardiomyopathy undergoing implantable cardioverter-defibrillator (ICD) implantation (n=20). Methamphetamine cardiomyopathy (MACMP) vs. Age- and sex-matched controls undergoing first ICD implantation without methamphetamine cardiomyopathy was evaluated on Defibrillation threshold (DFT) in Joules (J) during ICD implantation (p=<0.005). Methamphetamine cardiomyopathy was associated with higher defibrillation thresholds and a 50% requirement for high-output ICD devices compared to matched controls.
Synapse has enriched 10 related papers. Consider them for comparative context: