Key result
The Minix 8341 pacemaker had lower internal current (4.1 vs 16.1 microA) than the Pasys 8329, allowing programming from 5.0 to 2.5 V to extend projected longevity by 7.1 years versus 2.3 years.
Why the study?
Does the Minix 8341 pacemaker compared to the Pasys 8329 pacemaker improve battery current and projected longevity at lower pulse amplitudes?
Population
VVI pacemakers (Minix 8341 and Pasys 8329)
Comparison
Minix 8341 VVI pacemaker vs Pasys 8329 VVI pacemaker
Design
Other
Authors
Loading...
May extend projected longevity by enabling lower-voltage programming; leaves open real-world clinical validation in patients.
Observational
Does the Minix 8341 pacemaker compared to the Pasys 8329 pacemaker improve battery current and projected longevity at lower pulse amplitudes?
Programming pacemakers with low internal circuitry current to 2.5 V or less can significantly extend their projected longevity.
Schuchert et al. (1994) conducted an observational in Pacemaker longevity. Minix 8341 VVI pacemaker vs. Pasys 8329 VVI pacemaker was evaluated on Battery current and projected pacemaker longevity. The Minix 8341 pacemaker had lower internal current (4.1 vs 16.1 microA) than the Pasys 8329, allowing programming from 5.0 to 2.5 V to extend projected longevity by 7.1 years versus 2.3 years.