Key result
Titanium nitride-coated electrodes provided lower polarization signals and a higher functional rate of Autocapture compared to platinum helix electrodes.
Why the study?
Does endocardial electrode design affect polarization signals and the functional rate of the Autocapture algorithm in patients receiving pacemakers?
Observational (n=217)
Does endocardial electrode design affect polarization signals and the functional rate of the Autocapture algorithm in patients receiving pacemakers?
Titanium nitride-coated ventricular electrodes provide lower polarization signals, making them more suitable for pacemaker algorithms based on evoked response detection like Autocapture.
Answers are grounded in this paper and cite their sources.
Titanium nitride electrodes may improve Autocapture reliability; hypothesis-generating for lead selection but does not yet support practice change.
Some new generation pacemakers use an algorithm based on evoked response (ER) detection to verify beat-to-beat capture and to enable automatic adjustment of output. This is a prospective acute study of polarization signal (PS) and ER in nine currently available electrodes. Intraoperative testing of ventricular bipolar electrodes used the Autocapture (AC) algorithm. The intrinsic R wave, PS, ER, acceptance of AC function, and stimulation thresholds (STs) were obtained. Ventricular electrodes were categorized as follows: titanium nitride (TiN)-coated passive and active fixation, high impedance (HI), passive fixation (VP), iridium oxide-coated titanium (IROX) (VI), and platinum helix (PH) active fixation. Acute testing was performed in 217 patients with an average age of 74.26 years, 59.6% were men with primary pacing indication-SSS (46.3%). There were no significant differences found with respect to R wave and threshold between the various electrodes. PH active-fixation electrodes had significantly higher ER and PS than other groups including the TiN-coated active-fixation electrodes. TiN-coated electrodes (active and passive fixation) had significantly lower PS than other electrodes. As a result, TiN electrodes had a significantly higher functional rate of AC (91.7%), whereas PH had the lowest rate (0%). In conclusion, (1) polarization characteristics are significantly different for commercially available ventricular electrodes, (2) certain physical features at the tissue to electrode interface like TiN coating appears to be more important in determining PS than electrode tip size and fixation method, and (3) the current algorithm for AC requires electrodes that provide low polarization for satisfactory performance.
No takes yet. Share an insight, caveat, or question.
Lau et al. (2001) conducted an observational in Pacemaker indication (primarily Sick Sinus Syndrome) (n=217). Titanium nitride (TiN)-coated electrodes vs. Platinum helix (PH) and other electrode designs was evaluated on Functional rate of Autocapture (AC). Titanium nitride-coated electrodes provided lower polarization signals and a higher functional rate of Autocapture compared to platinum helix electrodes.