Key result
The peak endocardial acceleration sensor achieved a lower maximal heart rate compared to the activity sensor during treadmill exercise testing.
Why the study?
Does the PEA sensor provide adequate chronotropic response compared to an ACT sensor and healthy controls during exercise testing in pacemaker patients?
Cross-Sectional (n=29)
Does the PEA sensor provide adequate chronotropic response compared to an ACT sensor and healthy controls during exercise testing in pacemaker patients?
The PEA sensor functions hypochronotropically during exercise when programmed as a single sensor system, suggesting it should be combined with an activity-based sensor.
Answers are grounded in this paper and cite their sources.
PEA sensor associated with blunted peak heart rate versus activity sensor; leaves open optimal multisensor programming for chronotropic response.
There are limited data about the chronotropic capacity of the peak endocardial acceleration (PEA) sensor. This study directly compared the chronotropic function from the PEA and the activity (ACT) sensor. The study included 18 patients (age 73 +/- 7 years) with > or = 75% pacemaker-driven heart rate (HR) and a PEA sensor and 11 healthy controls (age 67 +/- 7 years) underwent a chronotropic assessment exercise protocol (CAEP) exercise test with the pacemaker patients in VVIR mode after programming the sensors in the default setting with adjustment of the upper sensor rate as an age related maximum value (220-age). The ACT sensor was externally strapped on the thorax. Achieved exercise duration for the patients and controls was, respectively, 9.2 +/- 3 vs 18.4 +/- 4 minutes (P <0.001). The maximal achieved HR with the PEA sensor was 124 +/- 25 beats/min, versus the ACT with 140 +/- 23, versus the controls with 153 +/- 26 beats/min (P <0.001 between the groups). For the PEA, ACT, and controls, the time to peak HR was, respectively, 11 +/- 3, 7 +/- 3.6, and 18 +/- 4 (P <0.001 between groups) and HR after 10 minutes recovery was, respectively, 80 +/- 20, 65 +/- 15, and 82 +/- 4 beats/min (P <0.001 between groups). The PEA sensor functions hypochonotroop during exercise programmed as a single sensor system. It is, therefore, preferable to combine the PEA sensor with an activity-based sensor in a dual sensor system. Although both groups had normal left ventricular functions, the exercise capacity of pacemaker patients is significantly lower than in the controls.
No takes yet. Share an insight, caveat, or question.
Erol-Yilmaz et al. (2004) conducted a cross-sectional in Pacemaker dependency (n=29). Peak endocardial acceleration (PEA) sensor vs. Activity (ACT) sensor was evaluated on Maximal achieved heart rate (beats/min) (p=<0.001). The peak endocardial acceleration sensor achieved a lower maximal heart rate compared to the activity sensor during treadmill exercise testing.
Synapse has enriched 10 related papers. Consider them for comparative context: