Key result
The CapSure Z lead reduced the mean current drained from the pacemaker compared to traditional leads, potentially increasing pacemaker longevity.
Why the study?
Does a very small surface area steroid-eluting pacing lead improve electrical performance and reduce current drain compared to traditional leads in patients receiving dual chamber pacemakers?
Cohort (n=188)
Does a very small surface area steroid-eluting pacing lead improve electrical performance and reduce current drain compared to traditional leads in patients receiving dual chamber pacemakers?
Answers are grounded in this paper and cite their sources.
The use of very small surface area steroid-eluting leads significantly reduces current drain, which combined with low energy output programming can increase pacemaker longevity.
We evaluated the handling performance at implant, and the long-term atrial and ventricular electrical performance of a new generation using a very small surface area (1.2 mm2) steroid-eluting electrode (Medtronic CapSure Z). We compared the performance of CapSure Z to that of traditional passive fixation leads, with and without steroid elution. The study was conducted during 2 years of follow-up. We studied 188 patients (105 males and 83 females; mean age 71 +/- 7 years). All of the patients were implanted with a dual chamber pacemaker and the same type of lead in both chambers. Forty-one patients received CapSure Z leads, 25 patients received Target Tip leads (8-mm2 surface area; no steroid elution), 63 patients received CapSure leads (8-mm2 surface area; steroid elution), and 59 patients received CapSure SP leads (5.8-mm2 surface area; steroid elution). The four groups were homogeneous in regards to sex, age, cardiac disease, and reason for implant. At follow-up, the CapSure Z lead showed sensing values comparable to the other leads, with lower pacing thresholds and higher pacing impedance in both chambers. We evaluated the mean current drained from the pacemaker by the different types of leads when using safe, low energy output settings. We found that by using CapSure Z leads, the mean current was significantly lower than that of the other types of leads (0.42 microA for CapSure Z ventricular lead vs 0.85 for CapSure SP, 1.42 for CapSure, and 1.54 for Target Tip). Thus, the use of the CapSure Z lead, combined with low energy output programming, will increase pacemaker longevity compared to the use of traditional leads and standard output programming.
No takes yet. Share an insight, caveat, or question.
MORACCHINI et al. (1999) conducted a cohort in Dual chamber pacemaker implantation (n=188). CapSure Z lead vs. Target Tip, CapSure, and CapSure SP leads was evaluated on Mean current drained from the pacemaker (ventricular lead). The CapSure Z lead reduced the mean current drained from the pacemaker compared to traditional leads, potentially increasing pacemaker longevity.