Key result
Patients with a DDD Genisis pacemaker and high threshold leads experienced a rapid fall in battery voltage from 2.7 to 2.49 V and a rise in impedance from 3 to 10.2 K ohms before replacement.
Why the study?
Does the use of high threshold leads alter end-of-life behavior in DDD Genisis pacemakers?
Observational (n=12)
Does the use of high threshold leads alter end-of-life behavior in DDD Genisis pacemakers?
High threshold leads combined with modern pacemakers with small batteries can cause rapid, unexpected falls in battery voltage near end-of-life, necessitating closer follow-up and reliance on surface ECG magnet rates for replacement timing.
Answers are grounded in this paper and cite their sources.
High-threshold leads may accelerate battery depletion in DDD Genisis pacemakers; leaves open whether magnet-rate monitoring improves replacement timing.
We noted a series of 12 consecutive patients with a DDD Genisis pacemaker that showed an unexpected and a relatively rapid fall in battery voltage and output as these devices approached end-of-life (EOL). Twenty-one of 24 leads were Vitatron Helifix leads and there was a relatively high mean threshold (atrial 2.5 +/- 0.94 V; ventricular 2.9 +/- 0.65 V). These devices were replaced after 65 +/- 12 months. During the 9.3 +/- 3.5 months before replacement, a striking fall in voltage from 2.7 +/- 0.04 V to 2.49 +/- 0.05 V was seen. Battery impedance rose from 3 +/- 1.2 K omega to 10.2 +/- 4.3 K omega during this same period. We unexpectedly observed a marked difference between programmed and telemetered output for both atrial (50%) and ventricular leads (30%). A discrepancy between measured and telemetered magnet rate was also seen. Despite this relatively rapid fall in battery voltage, several of these devices did not meet the manufacturer's recommended replacement time (RRT) criteria by magnet rate or according to the projected RRT determined by the relationship of battery impedance to current drain. These data have implications for the selection of RRT and EOL criteria for this device. Magnet rate measured by surface ECG was the safest indicator for RRT. Follow-up for this pulse generator should be increased to every 2 months when battery impedance is > 2 KOhms or if there is a difference between programmed and measured output amplitude of more than 15%. The data also highlight the effect of combining high threshold leads with modern pacemakers with relatively "small" batteries as well as certain problems with telemetered data.
No takes yet. Share an insight, caveat, or question.
Kraus et al. (1997) conducted an observational in Patients with a DDD Genisis pacemaker (n=12). DDD Genisis pacemaker with high threshold leads was evaluated on Fall in battery voltage and output. Patients with a DDD Genisis pacemaker and high threshold leads experienced a rapid fall in battery voltage from 2.7 to 2.49 V and a rise in impedance from 3 to 10.2 K ohms before replacement.