Key result
Implantation of a high impedance ventricular lead extended extrapolated pacemaker generator longevity compared to a standard impedance lead.
Why the study?
Does a bipolar high impedance ventricular lead improve estimated pacemaker generator longevity compared to a standard impedance lead in patients with a dual chamber pacemaker indication?
RCT (n=40)
randomized
Does a bipolar high impedance ventricular lead improve estimated pacemaker generator longevity compared to a standard impedance lead in patients with a dual chamber pacemaker indication?
Answers are grounded in this paper and cite their sources.
The use of high impedance ventricular leads significantly extends estimated pacemaker generator longevity compared to standard impedance leads without compromising pacing or sensing thresholds.
As pacemaker generator longevity is dependent on current consumption and resistance of the pacing lead, the use of a high impedance pacing lead theoretically results in an extension of battery longevity. Therefore, the effect of high versus standard impedance ventricular leads on generator longevity was studied. In 40 patients (21 women, age 73 +/- 13 years) with a standard dual chamber pacemaker indication, a bipolar standard impedance ventricular lead was implanted in 20 patients, the remaining patients received a bipolar high impedance lead in a randomized fashion. All patients received identical pacemaker generators and atrial leads. The estimated longevity of the generator was calculated automatically by a programmed pacemaker algorithm. After a mean follow-up of 39 +/- 4.8 months, no significant differences were observed with respect to mean pacing and sensing thresholds of the atrial and ventricular leads in both groups. However, the high impedance leads displayed a significantly higher impedance and a significantly lower current drain as compared to standard impedance leads (1,044 +/- 139 vs 585 +/- 90 Omega, and 2.2 +/- 0.4 vs 4.3 +/- 1.1 mA). The extrapolated generator longevity was significantly longer in the high impedance lead group, as compared to the standard impedance lead group (107.3 +/- 8.5 vs 97.6 +/- 9.0 months; P = 0.02). In conclusion, implantation of a high impedance lead for ventricular pacing results in a clinically relevant extension of generator longevity.
No takes yet. Share an insight, caveat, or question.
Berger et al. (2003) conducted an RCT in standard dual chamber pacemaker indication (n=40). Bipolar high impedance ventricular lead vs. Bipolar standard impedance ventricular lead was evaluated on Extrapolated generator longevity (p=0.02). Implantation of a high impedance ventricular lead extended extrapolated pacemaker generator longevity compared to a standard impedance lead.