Key result
High-impedance ventricular leads did not prolong actual pacemaker generator replacement time compared to standard impedance leads.
Why the study?
Does the use of high-impedance ventricular leads improve actual pacemaker generator longevity compared to standard impedance leads in patients with a dual-chamber pacemaker indication?
RCT (n=40)
randomized
Does the use of high-impedance ventricular leads improve actual pacemaker generator longevity compared to standard impedance leads in patients with a dual-chamber pacemaker indication?
High-impedance ventricular leads do not significantly prolong actual pacemaker generator longevity compared to standard impedance leads over long-term follow-up.
Answers are grounded in this paper and cite their sources.
High-impedance ventricular leads confer no generator longevity benefit; challenges assumptions from electrical properties and supports standard leads in dual-chamber pacing.
AIM: The long-term effects of high impedance vs. standard impedance pacing leads on actual generator longevity were studied. METHODS AND RESULTS: In 40 patients (21 females, age 73 +/- 13 years) with standard dual-chamber pacemaker indication, bipolar standard impedance ventricular leads and high-impedance leads were implanted in a randomized fashion. Identical pacemaker generators and atrial pacing leads were implanted in all patients. Patients were observed during a mean follow-up of 89.8 +/- 8.8 months before pacemaker replacement. Initially, the patients who received the high-impedance leads had a lower current drain as compared with standard pacing impedance leads, and the estimated pacemaker longevity was significantly prolonged, too. But this pattern disappeared after 6 years of follow-up, and finally the actual pacemaker generators' replacement time was 86.7 +/- 6.8 months in standard impedance lead group vs. 91.2 +/- 10.3 months in high-impedance lead group (P = 0.17). CONCLUSION: Implantation of high pacing impedance leads for ventricular stimulation does not result in a benefit with respect to pacemaker longevity as compared with standard impedance leads.
No takes yet. Share an insight, caveat, or question.
Etsadashvili et al. (2008) conducted an RCT in standard dual-chamber pacemaker indication (n=40). High-impedance ventricular leads vs. Standard impedance ventricular leads was evaluated on actual pacemaker generators' replacement time (p=0.17). High-impedance ventricular leads did not prolong actual pacemaker generator replacement time compared to standard impedance leads.