Key result
Returning to arc welding did not result in disturbance of sensing function or device inactivation in a patient with an implantable defibrillator following thorough screening.
Why the study?
Is it safe for a patient with an implantable defibrillator to return to arc welding?
Case Report (n=1)
Is it safe for a patient with an implantable defibrillator to return to arc welding?
A patient with an implantable defibrillator safely returned to arc welding after thorough screening showed no device inactivation or sensing disturbance despite electrogram artifacts.
Answers are grounded in this paper and cite their sources.
Supports case-by-case return to arc welding after screening in ICD patients; leaves open prospective validation for guidelines.
A male arc welder who was fitted with an implantable defibrillator wished to return to his former employment. Device testing and ambient magnetic field measurements were performed at his place of work. During welding, artifacts were seen on the intracardiac electrogram but there was no resulting disturbance of sensing function. Measured field strengths were too low to result in device inactivation. The patient resumed his work without incident. This appears to be the first reported case of a patient with an implantable defibrillator returning to this "electrically hostile" environment following thorough screening.
No takes yet. Share an insight, caveat, or question.
Embil et al. (1993) conducted a case report in Implantable defibrillator (n=1). Arc welding (ambient magnetic field exposure) was evaluated on Disturbance of sensing function or device inactivation. Returning to arc welding did not result in disturbance of sensing function or device inactivation in a patient with an implantable defibrillator following thorough screening.
Synapse has enriched 10 related papers. Consider them for comparative context: